Determination of As in ...

Published by

Determination of As in biological samples by ICP-MS is difficult for two reasons: the formation of the molecular ion 40Ar35Cl, which overlaps with monoisotopic As at a mass-to-charge ratio (m/z) of 75 (causing spectral interference), and signal enhancement due to organic matrix (nonspectral interference). Available procedures were examined, including the application of different correction procedures using 40Ar37Cl and 16O35Cl molecular-ion formation Dr Abdelsalam Asweisi